  "abstract": "Calibrate pH meters with at least two fresh buffers that bracket your readings (pH 7 then pH 4 for growing), before any session where decisions matter and at least weekly. Store glass pH probes wet in storage solution or pH 4/7 buffer, never in RO water or dry. Check EC meters against a 1.413 mS/cm standard. A pH probe with a slope below about 90-95% or one that can no longer be adjusted to buffer should be replaced; typical glass electrodes last one to two years.",